V1060示波器 您所在的位置:网站首页 V1565A V1060示波器

V1060示波器

2023-12-02 21:03| 来源: 网络整理| 查看: 265

X轴自动设业,CRT数字显示,Y轴灵敏度:2mv/div-5v/div.X轴扫速:50ns-0.5s/div,有X10扩展,示波管为15cm方屏内刻度。外形尺寸:280X130X350mm,重约:5.5kg ,220v/35w

The Hitachi portable read-out oscilloscope provides measure-ment reliability and ease of operation by employing amicroprocessor. The major features are: (1) Measurement information display The measurement information including the sweep speed, thedelay time, the IJNCAL display, and the voltage sensitivity ofthe vertical axis (V-1065 ,V-1065C and V-665 only) is alpha-numerically displayed on the CRT. Since troublesome setting operation procedures are eliminat-ed, an operator can concentrate on the displayed data formeasurement. (2) Measurement value display (V-1065, V-1065C, V-665 only) The distance between the two cursors displayed can read out the following; △V : Voltage between the reference cursor and the deltacursor △T : Time between the reference cursor and the deltacursor 1/△T: Reciprocal of the time The displayed data eliminates troublesome and time consumingcalculation procedures. Moreover, miscalculation of thescales is completely avoided. (3) Automatic time base range setting At a press of the AUTO button, an optimum time base range is automatically set.(A signal period from 1.6 to 4 cyclesapprox. is displayed.)The time base range is automatically changed to a correspond-ing change in input signal period.

(4) Trigger lock Since complicated pulse train waveforms are hard to trigger on, performing the trigger lock function enables sweep independent triggering. The "sweep time plus holdoff time"is fixed and a stable trigger is obtained at any time range.

(5) Bandwidth V-1065,V-1065C,: DC-100 MHz from 5 V/div to S mv/div V-1060  DC-20 MHz at 2 mV/div V-665, V-660 : DC-60 MHz from 5 V/div to S mv/div (6) High sensitivity DC-10 MHz at 2 mv/divHigh sensitivity of 2 mV/div is provided.

(7) Internal graticule Internal graticule lines eliminate parallax-viewing errorbetween the trace and the graticule lines. (8) Delayed sweep With delayed sweep, a portion of the signal can be magnifined for more accurate measurement and the windows of delayed sweep(B) is indicated on the main sweep(A) with cursors. (9)Auto trigger level Auto measuring of trigger level is employed, so that trigger level range is mathed to the trigger signal for maximum trigger sensitivity and stability. (10)TV triggering Exclusive TV sync separator circuit technology provides stable TV signal measurements on fields, frames and lines.

Specifications (A) Frequency counter 1.1 Specifications Measuring signal : Measure the signal selected as atrigger signal by the TRIG SOURCEOR X switch. Measuring range : 20 Hz to 100 MHz Accuracy of reference signal : 100 ppM (15 to 35°C)

Measuring range Display format Resolution Accuracy 20Hz≤f<100Hz 99.99Hz 0.01Hz Ref.signal ±1 LSD 100Hz≤f<1kHz 999.9Hz 1.0Hz Max 1kHz≤f<10kHz 9.999Hz 0.002Hz Max 10kHz≤f<100kHz 99.99Hz 0.04Hz Max 100kHz≤f<1MHz 999.9Hz 0.1Hz 1MHz≤f<10MHz 9.999Hz 0.002Hz Max 10MHz≤f<100MHz 99.99Hz 0.01Hz 100MHz≤f (999.9Hz) Not specified

1 .2 Frequency measurement   When MEASURE is selected by the SELECTOR switch, the AV cursor is displayed on CRT. Then press the SELECTOR switchthree times to display FREQ at the upper part of CRT.   The frequency counter is interlocked with the triggering andthe measurement starts from the triggering point.   Consequently, any special frequency having complex waveformscan be measured by changing the trigger level.   For example, when a signal as illustrated below is applied,the frequency of the entire waveforms can be measured whentriggered at point A. When triggered at point B, thefrequency of a pulse with large amplitude can be measured.

Note : When not triggered, the measurement can not be per-formed, since the counter is interlocked with thetriggering. In the non-triggering mode or in case ofthe input signal less than 20 Hz, the message"FREQ..." is displayed.

(B) Memory backup The panel setting data is retained for move than 48 hoursafter turning off power. It is not needed to perform thesame panel settings again even on the day after tomorrow.

(C) Other specifications Same with those of the V-1065.

V-1065/V-1065C V-1060 V-665 V-660 ●CRT Graticule 6-inch screen with internal graticule 0%, 10%, 90% and 100% markers 8x10 DIV(1DIV=1cm) Phosphor P31 Accelerating potential 17kV approx. 12kV approx. External intensity modulation Coupling: DC coupling Voltage: 5V or more Maximum input voltage: 30V (DC+AC peak)             or 30Vp-p AC at 1kHz or less Bandwidth: DC to 5MHz ●Vertical deflection system Sensitivity 2mV/DIV to 5V/DIV ±3%(switchable in 11 steps) Continuously variable Bandwidth DC to 100MHz -3dB 2mV/DIV: DC to 20MHz -3dB AC low pass: 10Hz 60MHz 10MHz Rise time 3.5 ns apporx. 2mV/DIV: 17.5 ns approx. 5.9ns approx. Delay time Leading edge can be monitoed 35ns approx. Maximum input voltage 400V(DC+AC peak)at 1 kHz or less Input coupling AC, DC, GND Input impedance 1M ohms ±1.5%, 23pF ±3pF Display modes CH1, CH2, DUAL, CHOP(250kHz approx.),ADD(DIFF mode can be established when the CH2 is in the INVERT mode.) Bandwidth limiting function 20MHz 10MHz Polarity selection ±(CH2 only) Common-mode rejection ratio 20dB minimum at 20MHz X-Y operation X-axis, Y-axis selectable Sensitivity X axis: CH1, CH2 2mV to 5V/DIV ±3% EXT      0.1V/DIV ±5% EXT÷10   1V/DIV ±5% Y axis: 2mV to 5V/DIV ±5% Phase error 3°or less from DC to 50 kHz X bandwidth DC to 500 kHz(-3dB) ●Horizontal deflection system Trigger mode Trigger,auto trigger Sweep mode Main sweep, continous delay sweep, alternate sweep, signal sweep Trigger source CH1, CH2, EXT(AC,DC,DC÷10),LINE TV trigger Excusive sync separtor circuit provided Sync polarity:- Trigger Sensitivity NORM mode Frequence DC to 20MHz 20MHz to 100MHz INT 0.35 DIV 1.5 DIV EXT 50 mV 150 mV DC to 10MHz 10MHz-60MHz AUTO mode Frequence 30 to 100Hz 100Hz to 20MHz 20 to 100MHz INT 1.5 DIV 1 DIV 1.5 DIV EXT 150 mV 100 mV 150 mV 30 to 100Hz 100Hz-10MHz 10-60MHz TV mode Sync signal Internal sync: 1div or more External sync: 200mVp-p or more Trigger level Variable range AUTO: Automatically corresponds to the trigger signal NORM: INT: ±4 DIV or more       EXT: ±0.4 V or more       EXT÷10:±4V or more External input impedance 1 M ohms ±5%, 25pF + 6pF External input voltage 400V(DC+AC peak) at 1kHz Sweep time A(main) sweep 50 ns/DIV to 0.5 s/DIV   Continuously variable (UNCAL) B(delay) sweep 50 ns/DIV to 50 ms/DIV Accuracy 10 to 35℃ 0 to 50℃ X1 ±3% ±4% MAG X10 ±4% ±6% Sweep magnification X10 Maximum sweep rate 5 ns/DIV Hold off time Variable Delay time 1μs to 5s Delay jitter 1/20,000 or less Alternate separation Variable Trigger lock function Provided Auto range function Provided Singnal sweep function Provided ●Readout funtion Pannel setting display Vertical axis: V/DIV(CH1,CH2),UNCAL,probe conversion Sweep speed: S/DIV,UNCAL,MAG(converted value) Other: Delay time Not provided Same as V-1065 Not provided ●Cursor readout function Voltage difference △V: △-REF Time difference △T: △-REF Frequency 1/△T: △-REF No cursor Same as V-1065 No cursor ●External output TRIGGER SIGNAL OUT Output voltage: 25mV/DIV approx. (Full scale on the CRT) 50-ohm termination Frequency response: DC to 10MHz Output impedance: 50 ohms approx. Not provided Same as V-1065 Not provided ●Calibrator Waveform 1kHz ±20%, square wave Voltage 0.5V ±1% ●Power supply 90V to 250V AC, 48 to 440Hz, 40W approx. ●Other Dimensions 275(W) x 130(H) x 360(D) mm Weigth 6kg EMI VDE0871, Category B 产品留言